BASIC
1.
Rotate the control knob to highlight the field on the right of BASIC.
2.
Press and rotate the control knob to select DOSE, INF RATE or DRIP RATE.
3.
The data in the trend table changes accordingly.
STEP
1.
Rotate the control knob to highlight the field on the right of STEP.
2.
Press and rotate the control knob to select a value in the range of 1-10.
3.
The data in the trend table changes accordingly.
DOSE TYPE
1.
Rotate the control knob to highlight the field on the right of DOSE TYPE.
2.
Press and rotate the control knob to select either DOSE/min, DOSE/hr, DOSE/kg/min or
DOSE/kg/hr in the popup menu.
3.
The data in the trend table changes accordingly.
UP-DOWN
1.
Rotate the control knob to highlight the UP-DOWN option in the window.
2.
Rotate the control knob to review more data.
REC
The REC option allows you to print the currently displayed data from the recorder.
NOTE
The titration table is independent from other functions of the monitor. The
patient information used in the titration table may not consist with the patient of
your monitor. Any change in the titration table will not affect the information of
the patient under monitoring.
10-4