11.5.5 Arrhythmia Alarm Setup
Select ARR ALARM >> in the ECG SETUP menu. The following menu appears. You can
change the settings of the arrhythmia alarm in this menu.
Figure 11-13 Arrhythmia Alarm Setup
In the menu, the ALM field indicates the alarm on/off status, REC indicates the alarm
recording on/off status and LEV indicates the alarm level. You can change the settings as
described below.
ALL ALM ON All alarms on
Enables all arrhythmia alarms;
ALL ALM OFF All alarms off
Disables all arrhythmia alarms that can be disabled;
ALL REC ON All recording on
Enables the recording of all arrhythmia alarms;
ALL REC OFF All recording off
Disables the recording of all arrhythmia alarms;
ALM LEV
Options: HIGH, MED and LOW.
Sets the level of all arrhythmia alarms that can be set to the
same value.
11-22