16.3 CO Setup Menu
Select the CO label in the CO parameter window, and the C.O. SELECT menu pops up. You
can select the C.O. SETUP option to enter the following menu.
Figure 16-6 C.O. Setup Menu
You can perform the following settings in this menu.
ALM TB alarm on/off status
ON: When a TB alarm occurs, the monitor gives alarm indications
and stores the alarm;
OFF: When a TB alarm occurs, the monitor neither gives alarm
indications nor stores the alarm;
When OFF is selected, the
icon is displayed on the right of the
TB label.
WARNING
The blood temperature alarm is disabled during CO measurement.
ALM LEV Alarm level
Options: HIGH, MED and LOW.
ALM REC Alarm recording
ON: When a TB alarm occurs, the monitor enables the
recording;
OFF: When a TB alarm occurs, the monitor does not enable
the recording.
16-8