2.
NURSE CALL
If this option is selected, the auxiliary output port will be able to output nurse call
signals, and you can perform the settings in the NURSE CALL SETUP submenu of
the USER MAINTAIN menu. For details, see 4.7.4 Nurse Call Setup.
If not selected, the nurse call function will be disabled and the NURSE CALL
SETUP>> option in USER MAINTAIN will become inactive. In this situation, you
cannot set the information in the NURSE CALL SETUP submenu.
3.
DEFIB. SYN (defibrillator synchronization signals)
If this option is selected, the auxiliary output port will be able to output defibrillator
synchronization signals. In this situation, you can turn on DEFIB SYNC in the
ECG SETUP menu to enable the defibrillator synchronization. For details, see 11.3
ECG Setup Menu.
If not selected, the defibrillator synchronization function will be disabled and the
DEFIB SYNC option in the ECG SETUP menu will set to OFF (it will not be
user-adjustable).