C.2.11 AG Module Alarm Messages
Alarm message A B L Cause Measure
AG SENSOR OFF No Yes 3 The AG sensor is
not properly
connected or is
disconnected.
Make sure the AG
sensor is properly
connected.
AG NO WATERTRAP No Yes 3 The AG water trap is
disconnected from
the monitor.
Make sure the water
trap is properly
connected.
CHANGE AG
WATERTRAP
Yes No 2 The AG water trap is
replaced.
Wait until the water
trap has been changed.
AG WATERTRAP TYPE
WRONG
Yes No 2 The AG water trap is
of a wrong type.
Replace with a water
trap of a correct type.
AG USA ERROR Yes No 2 AG module fault. Contact us to get rid of
the fault.
AG PARAMAGNETIC
O2 ERR
Yes No 2 O2 module fault. Contact us to get rid of
the fault.
AG GALVANIC O2
ERROR
Yes No 1
AG OXIMA DEPLETION
WARN
Yes No 2
AG OXIMA DEPLETION
ERR
Yes No 2
AG OCCLUSION Yes No 1 The actual pump rate
of the AG module is
<20ml/min, which
exceeds 1 second.
Remove the block of
the airway.
AG SELF-TEST ERR Yes No 1 AG module fails or
fails to communicate